NMS4110

Electrochemical Sensor Readout Integrated Circuit (ROIC)

Features

  • Level & time variant potential generation

  • Precise current measurement (~500pA)

  • 2/3-terminal electrochemical Sensor

  • 2-channel input (2 Working Electrode)

  • Chronoamperometry/Amperometric/Potentiometric

  • I2C/SPI interface

  • Sensor detection

  • Temperature sensor

  • Low power operation: periodic wakeup & shutdown

  • NVM for configuration storage

  • Working voltage: 2.5V ~ 3.3V

Block Diagram

Target Applications

Continuous Glucose Monitoring

  • Chronoamperometry for continuous 3-terminal (WE, RE, CE) glucose sensor

  • Chronoamperometry for continuous 2-terminal (WE, RE/CE) glucose sensor

 

Potentiostat for 2/3-terminal electrochemical sensor

  • Amperometric

  • Potentiometric

 

Biopotentiostat for 3-terminal electrochemical
Instrumental Amplifier

Applicable Sensor Configuration

Electrical Specification

  • TIA Gain: 1k ~ 500kΩ (internal), ~ 5MΩ (external)

  • CMRR of TIA: 210dB

  • DAC Resolution: 12bit or 6bit

  • DAC Range: ±800mV

  • DAC DNL: 0.496LSB (12bit) / 0.005LSB (6bit)

  • LPF Range: 47k ~ 4.7MHz (could be adjusted by external components)

  • DSM ADC: 16bit, ENoB=13bit

  • Temperature Sensor: 0.49°C